+86 185 3080 3982

Measurement & Analyzer Instrument

Home - Products - Measurement & Analyzer Instrument - Thermogravimetric analyzer combined with infrared spectrometer KS-MAI-TAIS850

Thermogravimetric analyzer combined with infrared spectrometer KS-MAI-TAIS850

The product combines the quantitative analysis capabilities of TG with the qualitative analysis capabilities of FTIR for a wide range of applications in the polymer, pharmaceutical and chemical industries.


Technical Parameters

Instrument introduction

 

Kunsheng has developed its thermogravimetric-infrared  combined system with its nearly 10 years of experience in R&D and manufacturing of Fourier transform infrared spectrometers, combined with the world's top thermogravimetric analyzer products. The product combines the quantitative analysis capabilities of TG with the qualitative analysis capabilities of FTIR for a wide range of applications in the polymer, pharmaceutical and chemical industries. The qualitative analysis of the gaseous products produced during the process of curing the cross linking reaction, substance decomposition or other reaction can be carried out.

 

Product features

(1) High quality, high cost performance

(2) Vacuum seal structure interface

(3) The system is equipped with an insulated heating transmission line and a special heated gas chamber for infrared spectrometer. The heating temperature can be set according to the need to effectively prevent the condensation of decomposition products. The special focusing light path design of the heated gas chamber enables it to achieve high sensitivity while achieving small volume.

(4) Infrared spectrometer with high sensitivity and high stability

(5) Infrared spectrometer adopts DLATGS, no liquid nitrogen is needed during the detection process.

(6) Infrared spectrometer can be adapted to different brands or models of thermogravimetric analyzer.


Technical parameters:

NameParameter
FTIR spectral range7800~350cm-1
FTIR resolution0.5cm-1
FTIR detectorImported high sensitivity DLATGS
Gas chamber length / volume100mm
Gas chamber designDouble cone light path, stainless steel
Gas chamber interior sizeMaximum inner diameter: 28mm; center inner diameter: 16mm
Gas chamber volume38.5ml
Infrared window size38 x 6mm
Transmission line size3mm OD x 1m
Transmission line materialTEFLON
exhaust pipe3mm OD TEFLON
Gas chamber maximum voltage24V 8A 200W
Transmission line maximum voltage24V 5A 100W
Gas chamber heating temperatureUp to 300 ° C
Transmission line heating temperatureUp to 230 ° C
Temperature control accuracy±1.0℃